ZHANG Mei, ZHOU Jun, YANG Li-dan, et al.Interference of Exogenous Insulin on Insulin Detection Test by Electrochemiluminescence Immunoassay.Journal of Sichuan University (Medical Science Edition),2018;49(6):924-928
Interference of Exogenous Insulin on Insulin Detection Test by Electrochemiluminescence Immunoassay
  
Key words:Interference test Exogenous insulin Electrochemiluminescence immunoassay
Author NameAffiliation
ZHANG Mei, ZHOU Jun, YANG Li-dan, et al 四川大学华西医院 实验医学科 
Hits: 1571
Download times: 0
Abstract:
      Objective To explore the interference of exogenous insulin therapy on insulin detection test by electrochemical luminescence immunoassay (ECLIA). Methods Insulin level was determined by ECLIA. According to the requirements of EP7-A2 of American Society for Clinical Laboratory Standards Institute Standards, paired difference experiment was conducted to evaluate the interference of 8 kinds of exogenous insulin on insulin detection, dose effect experiment was conducted to determine the relationship between exogenous insulin concentration and interference degree. Results When the interfering substance concentrations were ≤250 μU/mL, Gansulin N?, Gansulin R?, Humulin R?,Novolin R? and Lantus? all showed linear positive interference, while Levemir? showed a linear negative interference in high concentrations insulin and non-interfering in low concentrations insulin, Humalog? and Novo Rapid? showed non-interference in insulin detection. Conclusion The use of different exogenous insulin may have different interference on insulin measurement, which need laboratorians and physicians notice to avoid misdiagnosis.